CAUTION
This product must be connected to earth ground for proper operation. Failure to do so may result in
erratic operation.
3.6 ELECTRICAL INSTALLATION
All user supplied conduit ttings, junction boxes, etc. must be installed in compliance with
federal, state and local building codes.
3.6.1 Input Power Options
The F-4000 can be ordered with two different input voltage options. This is not user
selectable in the eld. The input power options are:
11.5 - 28.5 VDC, 10 Watts maximum
90 - 240 VAC 50/60 Hz, 15 VA maximum
WARNING
Conduit openings in the F-4000 enclosure must be closed with UL listed ttings applicable to NEMA
4 enclosures.
WARNING
The protective earth connection must be made as shown in Section 3.6.2. Failure to do so will result
in an increased risk of injury.
WARNING
All mains voltage connections must be made through the pre-drilled conduit/strain relief opening
located at the bottom of the enclosure. Failure to do so will result in an increased risk of injury.